MAM Phantoms

For acceptance and constancy tests for prone mammography stereotactic biopsy systems

  • For acceptance and constancy tests for prone mammography stereotactic biopsy systems
  • Designed in accordance with DIN 6868-163

 

The MAM technician (QRM-10148) and operator sets (QRM-10149) for NORMI® MAM digital are designed for acceptance and constancy tests for prone mammography stereotactic biopsy systems.  

Both sets offer the possibility to perform acceptance and constancy tests, according to DIN 6868-163. The phantoms are made of PMMA and can be positioned between the X-ray tube and detector of the mammography unit using two aluminium rods (included). 

  • Technician set for MAM 6868-163, QRM-10148  
  • Operator set for MAM 6868-163, QRM-10149  

Technical Specifications

Type Dimensions [mm] Quantity
Test plate 150 x 150 x 6 1 pcs.
Absorber plate 150 x 150 x 20 2 pcs.
Absorber plate 150 x 150 x 10 2 pcs.
Absorber plate 150 x 150 x 6 1 pcs.
Absorber plate 150 x 150 x 4 1 pcs.

All plates have two bore holes, such that they can be attached to the aluminum holder (two rods, diameter: 6 mm, length: 120 mm). 

Type Dimensions [mm] Quantity
Test plate 150 x 150 x 6 1 pcs.
Absorber plate 150 x 150 x 40 1 pcs.

All plates have two bore holes, such that they can be attached to the aluminum holder (two rods, diameter: 6 mm, length: 120 mm). 